Version 02.00, November 2014
Rohde & Schwarz R&S
®
BTC Broadcast Test Center 49
Analysis
Power measurements (R&S
®
BTC-K2055 option)
The power measurement requires the use of R&S
®
NRP-Zxx power sensors.
measurements depend on power sensor
measurement frequency coupled to
RF frequency
user-selectable measurement frequency
automatic setting of filter length
manual setting of filter length
depends on the R&S
®
NRP-Zxx power
sensor being used
R&S
®
NRP-Z11, R&S
®
NRP-Z21,
R&S
®
NRP-Z22, R&S
®
NRP-Z23,
R&S
®
NRP-Z24, R&S
®
NRP-Z51,
R&S
®
NRP-Z52, R&S
®
NRP-Z55,
R&S
®
NRP-Z91, R&S
®
NRP-Z92
BER measurements (R&S
®
BTC-K2060 option)
The bit error rate meter counts the errored bits in a data stream and displays the absolute error count and the relative error count
referenced to the total number of received bits. It is possible to analyze serial data streams or MPEG-2 transport streams.
At least one I/Q baseband generator (R&S
®
BTC-B1 option) is required for using the rear panel TS SERIAL IN or IP (1000BaseT)
interface.
for all digital modulation modes
depending on bit error rate
time window (1 s to 100 h), single run,
infinite
all measured values and error events
Inputs for BER clock, BER data,
BER enable
BNC female, USER 4, 5, 6 on rear panel
settable between 10 mV and 1.9 V
2
23
–1/2
15
–1 (in line with ITU-T O.151)
always, active high, active low
ASI, SMPTE 310, IP (see “Digital
audio/video interfaces”)
payload (PRBS in line with ITU-T O.151)
PRBS 2
23
–1/2
15
–1, constant 0x00/0xFF
NULL (1FFF (hex)), settable (0 to
1FFF(hex)), don’t care